Calibration, verification and adjustment are not the same thing
Three words that are used interchangeably on purchase orders but mean very different things. Knowing the difference helps you order the right service.
Purchase orders that arrive at the lab often ask for an instrument to be “calibrated and adjusted”, “verified”, or simply “calibrated”. These words are frequently treated as synonyms, but in metrology they describe three distinct activities. Ordering the wrong one can leave you with a certificate that does not meet your quality system’s needs.
Calibration
The International Vocabulary of Metrology (VIM) defines calibration as an operation that establishes the relationship between the values indicated by an instrument and the corresponding values provided by measurement standards, with their associated uncertainties.
In plain terms: we apply known values from a traceable standard, record what your instrument reads, and report the difference along with how confident we are in that difference.
Calibration on its own does not change the instrument and does not, strictly, say whether the instrument is “good”. It produces data.
Verification
Verification adds a judgement. It takes the calibration results and compares them with specified requirements, typically the manufacturer’s accuracy specification or a tolerance you supply, and states whether the instrument conforms.
Most customers want this. When an auditor asks “is this instrument fit for use?”, a verification statement gives a direct answer, provided the decision rule is clear.
Adjustment
Adjustment is a physical or software operation on the instrument that brings its readings closer to the reference. On modern instruments this is usually done through a password-protected calibration menu that stores new correction constants. On older equipment it may involve trimming potentiometers.
Adjustment changes the instrument. That is why as-found data must be recorded before adjustment takes place: once the constants are overwritten, the evidence of how the instrument was behaving during the last interval is gone.
Why the distinction matters
Consider a digital multimeter that arrives reading 0.02 % high on its 10 V DC range, against a specification of ±0.0035 %.
- A calibration reports the error and its uncertainty.
- A verification reports that the range is out of tolerance.
- An adjustment brings it back within tolerance, after which it is calibrated again to produce as-left data.
If you ordered only “adjustment”, you might receive an instrument that reads correctly but no record that it had been reading incorrectly for months. If you ordered only “calibration”, you would get accurate data showing a failure, but the instrument would come back still out of tolerance.
What we do by default
At PME Laboratory our standard service is calibration with verification against the manufacturer’s specification, with as-found data always recorded. If a point is out of tolerance, we contact you before adjusting, so the decision about how to proceed is yours.
If your quality system has specific requirements, such as tighter internal tolerances or no adjustment unless outside a defined limit, tell us when you request a quote and we will set up your account accordingly.