As our on-site service has grown, one of the most common questions we hear is whether a customer should send instruments to the lab or have us come to them. The honest answer is that it depends on the instruments, the quantity, and what matters most to your operation.

The case for in-lab calibration

Environmental control. Our lab is held at 23 °C ± 1 °C with controlled humidity and conditioned power. For precision instruments, such as 6½-digit and higher multimeters, reference standards and microwave analyzers, this control directly reduces uncertainty.

Full capability. Some standards are too large, delicate or numerous to transport. In the lab we have every reference standard available, so we can calibrate every function and range of an instrument, rather than a subset.

Adjustment and repair. If an instrument is out of tolerance, the lab has the equipment and time to adjust it properly, perform minor repairs, and re-calibrate.

Cost for small quantities. For a handful of instruments, sending them in is almost always cheaper than the cost of an engineer and equipment travelling to site.

The case for on-site calibration

Minimal downtime. Instruments that are part of a production line or test system can be calibrated in place, sometimes during a planned shutdown, without being removed and reinstalled.

Fixed installations. Some equipment simply cannot be moved: panel meters, installed test racks, large power analyzers on switchboards, or instruments that are part of a validated system.

System calibration. Calibrating an instrument together with its fixed cabling, fixtures and software can be more meaningful than calibrating it in isolation.

Logistics for large quantities. For dozens of instruments, the cost and risk of packing, shipping and tracking them may exceed the cost of an on-site visit.

Questions to ask

  1. How many instruments? As a rough guide, on-site work starts to make economic sense above about twenty instruments in one location.
  2. What accuracy do you need? If your requirements are close to the best capability of the instrument, the lab’s controlled environment may be necessary.
  3. Can the equipment be moved safely? Consider fragility, size and any requirement to re-validate a system after reinstallation.
  4. What does downtime cost? If an instrument being away for a week stops a line, on-site may pay for itself.
  5. Is there a suitable space? On-site calibration needs a stable, reasonably clean area with power, away from heavy vibration and draughts.

A hybrid approach

Many of our customers use both: precision bench instruments and reference standards come to the lab, while installed equipment and large quantities of handheld meters are calibrated on site. We record environmental conditions for on-site work on the certificate, just as we do in the lab, and adjust the uncertainty accordingly.

Start a calibration

Your next calibration is due sooner than you think.

Send us your instrument list and we will come back within one working day with a fixed quote, turnaround date and collection slot.