Preparing for an ISO 9001 audit: the calibration records auditors look for
Measurement equipment is one of the most common sources of audit findings. Here is what clause 7.1.5 asks for, and how to have the evidence ready.
If your organisation is certified to ISO 9001, measurement equipment is almost certainly on your auditor’s checklist. In our experience helping customers prepare, calibration is one of the most common sources of minor non-conformities, and almost all of them are avoidable with good records.
What the standard asks for
ISO 9001:2015 clause 7.1.5, Monitoring and measuring resources, requires organisations to determine and provide the resources needed to ensure valid and reliable results when monitoring or measuring is used to verify conformity of products and services.
Where measurement traceability is a requirement, or is considered essential to provide confidence in results, clause 7.1.5.2 requires that measuring equipment is:
- calibrated or verified, or both, at specified intervals or prior to use, against standards traceable to international or national standards;
- identified in order to determine its status;
- safeguarded from adjustments, damage or deterioration that would invalidate the calibration status.
It also requires the organisation to determine whether the validity of previous results has been adversely affected when equipment is found to be unfit for its intended purpose, and to take action as necessary.
The records that satisfy each requirement
An equipment register
A list of every instrument used for product or process verification, with a unique ID, description, location, owner, calibration interval and next due date. Auditors will often pick an instrument on the shop floor and ask to see it in the register.
Current certificates
For each instrument, the latest calibration certificate, showing results, uncertainties and traceability. The serial number on the certificate must match the instrument.
Status identification
A calibration label on the instrument showing the ID and due date, or another method of making status visible, such as a barcode linked to the register. Out-of-service or uncalibrated instruments should be clearly marked.
Acceptance criteria
Evidence that someone has reviewed each certificate and confirmed the instrument is suitable for its intended use. A simple signed and dated review stamp, or an entry in your register, is usually enough.
Out-of-tolerance actions
When an instrument is found out of tolerance, a record of the impact assessment: what was it used for, are any past results affected, and what action was taken? This is where as-found data becomes essential.
Protection from adjustment
Tamper seals over adjustment points or calibration menus, and procedures for handling and storage.
The findings we see most often
- Instruments in use that are past their due date.
- Certificates that do not match the instrument’s serial number.
- Certificates filed but never reviewed against the user’s requirements.
- No record of what happened after an instrument failed calibration.
- Personal instruments, such as an engineer’s own multimeter, being used for verification without being in the register.