Bench supplies and loads are often treated as utilities rather than instruments. If they set or measure values you rely on, they need calibrating too.

DC power supplies are on almost every electronics bench, and electronic loads are increasingly common in battery, power-electronics and production-test work. Because they are often thought of as sources of power rather than measuring instruments, they are frequently left out of calibration programmes. That is a mistake when their displays or programmed values are used to make decisions.

When does a power supply need calibrating?

Ask a simple question: do you rely on the number?

If a supply is only used to power a circuit while a separate, calibrated meter measures voltage and current, the supply’s own accuracy may not matter. If you set a voltage on the supply and assume that is what the circuit receives, or read the current from its display to calculate power consumption, then the supply is acting as a measuring instrument and should be calibrated.

Automated test systems are a particular case. Programmable supplies are often commanded to an exact voltage and their readback is logged as test data. In that situation, both the programming accuracy and the readback accuracy are part of the measurement.

What we check on a power supply

  • Voltage programming accuracy: the difference between the set voltage and the actual output, measured with a reference multimeter at several points across each range.
  • Voltage readback accuracy: the difference between the displayed or reported voltage and the actual output.
  • Current programming and readback accuracy: measured through a precision current shunt, with the supply operating into a suitable load.
  • Load regulation: how much the output voltage changes between no load and full load.
  • Line regulation: how much the output changes with variations in mains supply voltage, where specified.
  • Ripple and noise: measured with an appropriate bandwidth limit, commonly 20 MHz, and reported as RMS and peak-to-peak.
  • Over-voltage and over-current protection trip points, on request.

What we check on an electronic load

Electronic loads are checked in each operating mode they offer:

  • Constant current (CC) setting and readback accuracy
  • Constant voltage (CV) setting and readback accuracy
  • Constant resistance (CR) and constant power (CP) accuracy
  • Voltage and current measurement accuracy

For high-current loads, we use precision shunts and a high-capacity source, and allow time for thermal stabilisation, because shunt resistance changes with temperature.

Remote sensing

Many supplies offer remote sense terminals, which compensate for voltage drop in the load leads. If you use remote sensing in your application, tell us, and we will calibrate with it connected. The difference between local and remote sense performance can be significant at high currents.

Practical tip

If you have a supply that is only ever used as a power source, label it clearly as “Indication only – not for measurement” and record it in your equipment register as not requiring calibration. Auditors are generally comfortable with that, provided the label and the register agree and the supply is not actually being used for measurement.