Heat, humidity and dust: how the UAE summer affects your test equipment

Instruments are specified for a comfortable lab at 23 °C. Many of ours spend August in a vehicle on a site in Ras Al Khaimah. Here is what that does to them.

It is the end of August, and anyone who has worked outdoors in the northern Emirates this month knows what the climate can do to people. It does similar things to instruments. Every summer we see a rise in failures that trace back not to age or misuse, but to heat, humidity and dust.

What the specification assumes

Most test equipment is specified at a reference temperature, commonly 23 °C ± 5 °C, with a temperature coefficient that tells you how much extra error to add for every degree outside that band. Operating ranges for handheld instruments often extend to 50 °C, and storage ranges may be higher, but performance outside the reference band is always worse, and the specification will tell you by how much.

A vehicle parked in the sun in July can easily exceed 60 °C inside. That is often beyond the storage rating of the instrument, never mind its operating range.

Heat

Sustained high temperature accelerates the ageing of electronic components. Voltage references drift, electrolytic capacitors dry out, LCD displays fade, and batteries lose capacity and can swell. In our experience, handheld meters stored in vehicles over the summer are noticeably more likely to be found out of tolerance at their next calibration than identical meters kept indoors.

Humidity and condensation

The coast of Ras Al Khaimah combines high humidity with heavily air-conditioned buildings. Moving an instrument from a cold office to a humid outdoor environment causes condensation on and inside it within seconds. Moisture on high-impedance circuits causes leakage currents, which show up as errors on high-resistance ranges and on insulation testers in particular. Repeated cycles lead to corrosion of contacts and PCB traces.

Dust and sand

Fine dust gets into rotary switches, input jacks, cooling fans and connector interfaces. On RF instruments, a contaminated connector can cause measurable reflection and loss, and grit trapped in a precision connector can permanently damage both it and whatever it is mated with.

Practical advice

  • Do not store instruments in vehicles. If they must travel, use an insulated case and take them indoors at the end of the day.
  • Let instruments acclimatise. After moving between environments, give an instrument time to reach ambient temperature before making precision measurements. If condensation is visible, wait until it has fully evaporated before switching on.
  • Use desiccant in storage cases and replace or regenerate it regularly.
  • Cap RF connectors whenever they are not in use, and clean them with appropriate tools and isopropyl alcohol.
  • Check fans and filters on bench instruments. A blocked filter can push internal temperatures well beyond design limits.
  • Remove batteries from instruments that will be stored for long periods.

When to consider a shorter interval

If an instrument spends most of its life on outdoor sites, its calibration interval should reflect that. We are always happy to look at the history of your field instruments and recommend whether a shorter interval would be justified.